Performing the measurement
MAGNETOM Sempra
Print No. M4-08006G.623.01.02.02
7
If necessary, tilt the coil slightly in the direction of the magnet
(with the tilt support mechanism) to position the patient more
comfortably.
1
Align the laser light localizer with the center marking of the coil.
2
Start the measurement.
Quality measurement
4.11.3
✓
The Spine 18 or the table pads are positioned on the patient table.
✓
Lower part of the Foot/Ankle 10 is positioned on the patient table.
1
Position the Foot/Ankle phantom correctly in the lower part of the
coil.
2
Place the upper part of the coil onto the base plate.
3
Connect the coil plug to coil socket 2 or 1 on the patient table.
4
Align the laser light localizer with the center marking of the coil.
5
Start the quality measurement
assurance (software)).
4.12
Flex Coil Interface
Description
4.12.1
The Flex Coil Interface is required for connecting the Flex coils.
Coils
(
Page 33 Performing quality
4