Installing The Monitor; Bottom View Of Monitor - Siemens SONOLINE G50 Service Manual

Ultrasound systems
Hide thumbs Also See for SONOLINE G50:
Table of Contents

Advertisement

Chapter 4 - Removing and Replacing Spare Parts
17. Place the defective monitor in the packaging and prepare it for return.

2.3 Installing the Monitor

1. Before mounting the replacement monitor, verify that all cables are routed
!
2. Use the Allen wrench provided to install the fasteners as illustrated in
Warning: The improper installation of the threaded fasteners that support the monitor can cre-
!
ate a potential hazard. The monitor is heavy and could cause serious injury. Always verify that
the two cap screws (M4 X 18) and one bolt are tightened securely so that the monitor will not
loosen and fall.
3. Attach the bottom cover to the monitor (Figure 4-6) using the two M4 X 18
4. Verify that no cables are pinched by the monitor installation (Figure 4-5).
Warning: The safety ground for the G50 and the G60 S is shown in Figure 4-5. Improper instal-
!
lation of this cable can produce shock hazards that will cause serious injury or death to the
operator, to the CSE, and to the patient. Always verify that the safety ground is properly con-
nected and securely fastened. Insert the M4 X 12 screw through the lug attached to the safety
ground cable, place the external tooth M4 lock washer on the screw, between the lug and the
chassis, and install the screw in the chassis.
Siemens
Figure 4-7 Bottom View of Monitor
outside of the monitor post toward the rear of the G50 or the G60 S frame.
Caution: Use fasteners of the correct length to prevent damage to underlying
equipment. See "Fastener Reference" (page H-1).
Figure 4-6 (top right photo): Tighten the two screws (M4 X 18) and one bolt.
Note: When attaching the bottom cover, ensure that the small hook (Figure 4-6)
of the monitor is lodged in the square hole of the bottom cover.
screws.
7482255 Rev 01
October 2002
Page 7 of 52