Installing or Updating the System Software
4. If the DIMAQ-IP software will be updated, back up all exam and patient
!
5. If deemed necessary, uninstall the feature keys from the system according
6. If the system software is being updated to version 1.2, update the DIMAQ-
7. Insert the MO diskette containing the version 1.2 software in the MO drive.
8. Press the F4 key to display the Preset screen shown in Figure 8-1.
Page 4 of 42
data to a CD before performing the DIMAQ-IP software update.
Caution: If the system software will be upgraded (or downgraded to an earlier
version) for testing or to resolve an operational problem, the feature keys for
options that are not supported by the installed version will be lost. To avoid this
problem, uninstall all feature keys before upgrading or downgrading the system
software. See "Uninstalling the Product Feature Keys" (page 8-14).
The uninstallation of all feature keys is a precaution. Experienced customer ser-
vice engineers (CSEs) may choose to leave the keys installed.
to the instructions provided in "Uninstalling the Product Feature Keys"
(page 8-14).
IP software to version 1.2 before continuing this procedure: See "Install-
ing or Updating the DIMAQ-IP Software" (page 8-7).
Note: The version 1.2 software update is described in "Installing the Version 1.2
Software Update on the G60 S" (page 7-1)
.
Scroll down to System/Language menu item.
Figure 8-1 Preset Screen
7482255 Rev 01