Chapter 7 - Installing the Version 1.2 Software Update on the G60 S
15. Select the DIMAQ-IP Maintenance button located on the Main Menu
16. Insert the CD containing the version 1.2 DIMAQ-IP software, in the CD
17. Move the mouse pointer to the bottom of the screen to display the Windows
18. Select Shutdown → Restart from the menu.
19. The DIMAQ-IP will reboot from the software CD; the display of the BIOS
20. The blue Radisys logo will be displayed.
21. The software installation requires approximately 5 minutes: The R:\>
22. After the R:\> prompt appears, remove the software CD.
23. Operate the power switch to shut down the system: After two minutes, if the
24. After the system has powered down, wait 15 seconds and re-energize the
25. Press the F4 key and select the Service menu item.
26. Enter the service password.
27. Select the DIMAQ-IP Maintenance button located on the Main Menu
28. Enter the Windows 2000 Product Identification code and select the Ok but-
29. Wait until the DIMAQ-IP powers down, then de-energize the system.
30. Wait 15 seconds and re-energize the system.
31. Allow the DIMAQ-IP to boot completely; the boot process requires approx-
32. Press the Review button.
33. Observe the DIMAQ Study Screen with the HD radio button selected; this
34. Create an image file, using any transducer, and assign a patient name and
35. Export the image file to the CD to verify its operation.
36. Verify that the exported image can be viewed by the system.
37. Delete the exported image from the HDD.
Siemens
screen to display the Windows 2000 screen running on the DIMAQ-IP mod-
ule.
drive.
toolbar, and select the Start button.
and Windows boot screens will be followed by the software installation win-
dow.
prompt appears when the installation is completed.
system has not powered down, operate the circuit breaker located at the
rear of the system to de-energize it.
system and allow it to boot.
screen to display the Windows 2000 screen running on the DIMAQ-IP mod-
ule.
ton.
imately 5 minutes.
indicates the success of the software update.
patient identification number to it; this file will be used to verify the operation
of the CD.
7482255 Rev 01
October 2002
Page 5 of 8