Siemens SONOLINE G50 Service Manual page 143

Ultrasound systems
Hide thumbs Also See for SONOLINE G50:
Table of Contents

Advertisement

Chapter 8 - Software
!
Siemens
f. In the File Type menu, select General System Presets.
g. Select the Store button.
h. Enter a file name using the keyboard: Include the date and data
description in the file name (MMDDYYGenSysPre).
i.
Select the Ok button; select the Ok button to close the message box
and to continue.
j.
In the File Type menu, select OB Tables and Formulas.
k. Select the Store button.
l.
Enter a file name using the keyboard: Include the date and data
description in the file name (MMDDYYOBTabFor).
m. Select the Ok button; select the Ok button to close the message box
and to continue.
n. In the File Type menu, select Default Settings.
o. Select the Store button.
p. Enter a file name using the keyboard: Include the date and data
description in the file name (MMDDYYDefSet).
q. Select the Ok button; select the Ok button to close the message box
and to continue.
r. In the File Type menu, select Total System Configuration.
s. Select the Store button.
t. Enter a file name using the keyboard: Include the date and data
description in the file name (MMDDYYTotSysCon).
u. Select the Ok button; select the Ok button to close the message box
and to continue.
v. Select the Utility tab and double-click MO and SONOLINE to view the
contents of the MO diskette.
w. Verify that the stored files appear on the diskette: Service Data Backup,
MMDDYYGenSysPre, MMDDYYOBTabFor, MMDDYYDefSet, and
MMDDYYTotSysCon.
x. Select Exit.
y. Eject and store the MO diskette.
Caution: Update of the DIMAQ-IP software will destroy existing exam and
patient information; back up exam and patient data before updating the DIMA-IP
software.
7482255 Rev 01
October 2002
Page 3 of 42

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sonoline g60

Table of Contents