Chapter 4 - Removing and Replacing Spare Parts
1. Turn off the system power and disconnect the power cord from the outlet.
Warning: The G50 and the G60 S contain hazardous voltages that can range from 90 Vac to
!
264 Vac. These voltages can cause serious injury or death. Only authorized service personnel
are permitted to remove protective covers and access the energized components of this sys-
tem.
2. Loosen the two screws located at the bottom of the front cover, and pull out
The screws designated as CK3x8 in Figure 4-13 are M3x8 binding head screws with exte
nal lock washers. The CK4x8 screws are M4x8 binding head screws with external lock
washers. The screws designated as S3x8 are M3x8 flat head screws. See "Fastener Refe
ence" (page H-1).
Siemens
the top edge of the cover as shown in Figure 4-13.
Figure 4-13 Removing the Front Cover
7482255 Rev 01
October 2002
Page 15 of 52