3.2.1 Temperature Specifications

The G50 and the G60 S can be operated in the environment listed in Table 1-5.
!
Specification
System Operating Temperature
System Transport Temperature
Array Transducer Operating Temperature
Array Transducer Transport Temperature
Mechanical Sector Transducer Operating Temperature
Mechanical Sector Transducer Transport Temperature
Pencil CW Transducer Operating Temperature
Pencil CW Transducer Transport Temperature
(a)
Operation of the MO drive is guaranteed to 35

3.2.2 Humidity Specifications

The G50 and the G60 S are designed to operate in an environment that can
experience humidity conditions ranging from 30% to 85% relative humidity,
non-condensing. During storage and transport, the G50 and the G60 S can
withstand from 0% to 95% relative humidity, non-condensing.

3.2.3 Altitude

The G50 and the G60 S design can operate at altitudes from 1060 hPa (-400 m
or -1312 feet) to 700 hPa (3050 m or 10,000 feet).
Siemens
Caution: The G50 and the G60 S and the installed peripherals (OEMs) contain
cooling fans that are required to maintain the equipment at acceptable operating
temperatures. Over-temperature conditions can impair the accuracy of the sys-
tem and damage its components. Do not attempt to operate the G50, the G60 S,
or the installed peripherals if any fan becomes inoperable.
See "Diagnostic Indicators" (page 6-9): LED 8 illuminates when the temperature
in the E module exceeds the over-temperature limit of 80 °C (176 °F). Consult
the Instructions for Use for information on the warning message displayed when
operating temperatures are exceeded.
Table 1-5 Temperature Specifications
(a)
7482255 Rev 01
Value
10 °C to 40 °C
-10 °C to 55 °C
10 °C to 40 °C
-10 °C to 60 °C
20 °C to 40 °C
-5 °C to 50 °C
10 °C to 40 °C
-10 °C to 60 °C
°C
°F
(95
)
October 2002
(50 °F to 104 °F)
(14 °F to 131 °F)
(50 °F to 104 °F)
(14 °F to 140 °F)
(68 °F to 104 °F)
(23 °F to 122 °F)
(50 °F to 104 °F)
(14 °F to 140 °F)