6.8 Diagnostics Screen

Access the Diagnostics screen shown in Figure 8-19 by selecting the
Diagnostics button located on the Main Menu screen (Figure 8-10).
On the Diagnostics screen, check boxes arranged according to hierarchy
levels are displayed adjacent to related test targets. Checking the box adjacent
to the All Packages label will enable testing of all packages. Checking a box in
the next lower hierarchy will remove the check mark from the All Packages
target and enable testing of all of the targets in the hierarchy that is located
below that checked target. Placing a check in a single box located at the lowest
hierarchy will cause only that target to be tested.
When check marks have been entered in the boxes for the targets that are to
be tested, select the Execution button to test the selected targets. When the
pattern check for the host module, E module, or the DIMAQ-IP is performed,
power down the system and re-energize it to reboot the software.
Observe the monitor as the testing progresses; pass–fail results will be
displayed in the window located to the right of the diagnostic list.
See "Diagnostic Messages" (page 6-3) for information on the messages. To
determine which spare part contains a failed component cited in the diagnostic
message, see "Component Designations" (page F-1); "Spare Parts" (page C-
1) lists the spare parts.
Siemens
Service
Diagnostics
Site
All Packages
System
HW / SW
Diagnostics
Log
Service History
Backup / Restore
Figure 8-19 Diagnostics Screen
7482255 Rev 01
October 2002
HostBox Fail:
Host Module Package
Pattern Check
Host Module
E Module Package
Pattern Check
E Module
IPPC Pattern Check
Operator
Power Box
Home
Test #1 Erro
Execution
Page 29 of 42

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sonoline g60

Table of Contents