Installing the G50 Third Array Port
5. Detach all transducers from the G50.
6. As depicted in Figure 5-11, loosen the two screws located at the bottom of
7. Pull the top of the front cover away from the G50 to release the three pairs
8. Grasp the top of the front cover and one of the bumpers; pull and detach
9. Remove the two labels indicated in Figure 5-12 from the front of the probe
10. Discard the two ("Do not open") labels; they will not be reinstalled.
11. As indicated in Figure 5-12, remove the 12 screws (M4 X 8); retain these
12. After removing the 12 screws, pull the plastic loop (Figure 5-12) toward the
13. From the rear of the probe interface panel just removed, unscrew the two
Page 12 of 16
the front cover of the G50.
of magic fasteners (Figure 5-11).
the front cover from the G50.
Figure 5-11 Removing the Front Cover
interface panel; these labels carry the message, "Do not open."
screws for reinstallation during the reassembly of the G50.
front of the G50 to remove the probe interface panel.
screws to release the blank panel indicated in Figure 5-12.
Main unit
Magic fastener (3)
Front cover
Screw (2)
Bumper
7482255 Rev 01