Chapter 4 - Removing and Replacing Spare Parts
2. Position the host module on the rear side of the G50 or the G60 S.
!
!
3. Fasten the host module by using the four screws removed previously.
4. Secure the rear cover onto the rear side of the G50 or the G60 S: Use the
5. When connecting the cables to the host module, verify each connector

9.4 Confirming the Operation of the Host Module

1. Connect the system power cable to a properly grounded outlet.
2. Connect a transducer to the G50 or the G60 S.
3. Turn the power switch On.
4. Turn the Standby switch On.
5. Confirm that a B-mode image is displayed on the monitor.
6. Press the F4 key and select the Service software.
7. Activate the Service software according to the instructions provided in "Ser-
8. Access the "System Information Screen" (page 8-24).
9. Verify the system serial number and the system software version.
Siemens
Caution: Cables can be pinched and damaged when the host module is
installed. Verify that no cables are pinched between the module and the edge of
the OEM interface panel or the LED indicator panel.
Caution: Use fasteners of the correct length to prevent damage to underlying
equipment. See "Fastener Reference" (page H-1).
procedures provided in "Removing the Host Module" (page 4-41), in
reverse order to install the cover.
number to prevent mis-connection that could damage the equipment.
vice Software Operation" (page 8-17).
7482255 Rev 01
October 2002
Page 43 of 52