Chapter 4 - Removing and Replacing Spare Parts
6. Release the cables from the cable clamp located at the upper left of the
7. Remove the cable clamp located at the right side of the DIMAQ-IP; this
8. Disconnect the three cables from DIMAQ-IP module: When disconnecting
9. Disconnect the remaining cables from the DIMAQ-IP module.
10. Loosen the three thumbscrews of the DIMAQ-IP module shown in
!
11. Create sufficient clearance for dismounting the DIMAQ-IP module safely by
12. Move the DIMAQ-IP module to the left and dismount it from the system.
Siemens
DIMAQ-IP module.
Tie-wrap these cables to the adjacent panel.
Figure 4-20 DIMAQ-IP Module Installation
cable clamp is shown on the right side in Figure 4-20 and requires the
removal of the screw.
the DIMAQ-IP power cable, unscrew its cable clamp as one piece
(Figure 4-20).
Figure 4-21, using the short number 2 Phillips screwdriver.
Caution: The close clearances around the DIMAQ-IP create the potential for
cable damage. Verify that no cable is pinched or abraded when the DIMAQ-IP is
removed.
routing the cables in front of it, and to the sides of the DIMAQ-IP module.
DIMAQ-IP
7482255 Rev 01
October 2002
Page 25 of 52