Removing and Installing the DIMAQ-IP Module

5.3 Installing the DIMAQ-IP Module

To install DIMAQ-IP module on the G60 S, perform the reverse of the
procedures described in "Removing the DIMAQ-IP Module" (page 4-24). See
"Installing the G50 DIMAQ-IP" (page 5-2) for installation instructions describing
the installation of the DIMAQ-IP option on the G50.
!
!

5.4 Confirming the Operation of the DIMAQ-IP Module

1. Connect the system power cable to a properly grounded outlet.
2. Connect a transducer to the system.
3. Turn the system power switch On.
4. Turn the Standby switch On.
5. Confirm that an image is displayed on the monitor.
6. Press REVIEW.
7. Confirm that the monitor shows the DIMAQ-IP module (Digital Lab) screen.
Page 26 of 52
DIMAQ-IP
Figure 4-21 DIMAQ-IP Module Thumbscrews
Caution: When installing the DIMAQ-IP module, verify that no cables are
pinched between parts or surfaces. Push the DIMAQ-IP module in slowly, veri-
fying that no cables are pinched between the DIMAQ-IP module and the sur-
rounding equipment.
When connecting cables between the DIMAQ-IP module and the host module,
read the connector labels carefully and confirm that the cables are attached to
the proper connector.
Caution: Use fasteners of the correct length to prevent damage to underlying
equipment. See "Fastener Reference" (page H-1).
7482255 Rev 01
October 2002
Thumbscrews (3)
LAN Host to P33
LAN Net to P30
Siemens