Removing and Installing the MO Drive
14. Move to the front side of the G50 or the G60 S and pull off the sub control
15. Press and release the hooks on both sides of the ribbon cable connector,
16. Hold the MO drive with your hand, inserting one hand from the rear side of
!
Page 38 of 52
panel.
and disconnect the cable from CN2 on the reverse of the sub control panel.
Figure 4-32 Removing the Monitor Post Cover
Figure 4-33 Removing the Sub Control Panel (LCD)
the G50 or the G60 S, while removing the screws from the MO drive bay.
Caution: When the four screws are removed from the MO drive bay, the bay can
fall free. Use caution, holding the MO drive bay securely, to prevent if from drop-
ping and being damaged.