Chapter 9 - Planned Maintenance

4.0 Preparations

Perform the following preparations before beginning the maintenance routines.
Warning: When handling parts that may have had patient contact, take appropriate precau-
!
tions against exposure to blood borne pathogens.
1. Discuss the system performance with the customer; determine what con-
2. In the maintenance report, record the information obtained from the cus-
3. Save the system preset data on an MO diskette.
4. Turn off the system and disconnect the power cable from the external

5.0 Cleaning the System

Use a soft cloth, lightly moistened with a mild soap solution, to clean the
surface of the system, including the system control panel, transducer holders,
system side panels, and any on-board OEM peripheral devices. Refer to the
operating instructions for additional information on cleaning the system,
transducers, and filters.
!
6.0 Cleaning Subsystems
Clean the following subsystems:
1. On each side of the system, remove the air filter retainer (2 total) as indi-
Siemens
cerns the customer might have regarding the system performance.
tomer, and the actions taken to resolve these concerns.
power source.
Do not use alcohol, other organic solvents, or cleansing agents
Caution:
that contain alcohol or organic solvents to clean the
cated in Figure 9-1.
Figure 9-1 Air Filter Removal
7482255 Rev 01
October 2002
.
G60 S
Page 3 of 8