32
Tableside control panel (NBF), board D41
31 23 879 J1042,
Board D41 is coded in the factory to "Patient table basic unit".
Recoding for use on the control console, board D41
•
Place the tableside control panel with the membrane keyboard side on a level surface.
•
Remove four screws (M 2.5x16) from the lower part of the housing.
•
Carefully lift off the lower housing.
Four metal spacers are lying loosely on the PC board.
•
Set or check the DUEP slave address of switch S3 according to the application and the
wiring diagram:
Fig. 2:
Assembly
•
Center the four metal spacers on the contact surfaces and insert the four screws
(M2.5x16) in the holes in the lower part of the housing.
•
Join the housing parts together so that one of the screws can be inserted through the
hole of the grounding tape.
The metal spacers must not move during assembly.
•
Tighten the screws.
LITHOSTAR Multiline
S3 Switch
RXL2-120.841.02.01.02
09.05
PC Board Replacement
wiring diagram J1042-61
Address 7: Tableside control panel
(NBF) basic unit (patient table) factory
setting
Address 2: Tableside control panel
(NBF) basic unit control console
Page 32 of 72
CS PS 24
0
0
0
Siemens AG
Medical Solutions