Replacing The Oem Outlet Fuse; Replacing The Fuses; Xdst A07-1 (07473122) - Siemens SONOLINE G50 Service Manual

Ultrasound systems
Hide thumbs Also See for SONOLINE G50:
Table of Contents

Advertisement

Replacing the OEM Outlet Fuse

12.0 Replacing the OEM Outlet Fuse
Use the instructions provided in this section to remove and install the power
fuse located in the host module. Replacement of the fuse requires
approximately 5 minutes.
!
12.1 Tools Required
Use a number 2 Phillips screwdriver to perform this procedure.

12.2 Replacing the Fuses

1. Turn off the system and unplug the power cable from the wall outlet.
Warning: The G50 and the G60 S contain hazardous voltages that can range from 90 Vac to
!
264 Vac. These voltages can cause serious injury or death. Only authorized service personnel
are permitted to remove protective covers and access the energized components of this sys-
tem.
2. Loosen the two captive screws from the upper side of the rear cover and
3. Loosen the two captive screws from the lower rear panel and remove the
4. Pull the upper side of the rear cover toward you.
5. Locate the XDST (A07-1) PC board on the right side of the host module;
Page 50 of 52
Caution: An electrical failure within the system causes the power fuse(s) to
open. Before replacing the power fuse, discover and correct the failure that
caused the fuse to open.
remove the cover.
panel.
see Figure 4-44.
Figure 4-44 XDST A07-1 (07473122)