Chapter 6 - Troubleshooting
B-mode image
Probe select
adjust Ok?
CINE mode
All B-mode
Siemens
C
Perform loop test 4.
Service >
Diagnostics >
No
E Module Package >
Ok?
Pattern Check.
Recycle power.
Loop test 4 and
3 both Ok?
Yes
No
G60 S System
Ok?
Yes
TX power
Test; Service >
No
Diagnostics >
Power Module.
Yes
After test pattern
check, recycle the
system power
Ok?
Yes
problems
resolved?
Yes
Return
Perform loop test 3.
Service >
Diagnostic >
Power Module.
No
No;
4 pass
3 fail
Failed loop test 4;
passed loop test 3.
Yes;
4 fail
3 pass
Yes
Third Array
No
Port Option
Installed
Yes
Remove Third
Array Port and
Problem
Retest
Remains
Yes
Yes
Power module
diagnostic
passed.
Yes
No
No
Figure 6-7 Chart C: B-mode Imaging
7482255 Rev 01
October 2002
Replace host module
Replace cable
Replace E module
No
Replace G50 LC E module
Replace G50 Third Array Port
No
Replace G50 LC E module
Replace G60 S E module.
No
Replace power module.
Replace E module.
Replace E module.
Replace E module.
Page 17 of 22