Removing and Installing the Control Panel
6.2.2 Removing the Sub Control Panel (LCD)
1. From the front side of the system, pull the sub control panel forward and
2. On the reverse side of the sub control panel, press the detents on both
3. Remove the sub control panel from the upper G50 or the G60 S.

6.2.3 Removing the Main Control Panel

1. On the control panel, push down on the detent (Figure 4-25) that holds
2. The screws located at the sides of the control panel are now visible:
3. Remove the two screws from the center part of the front handle, indicated
4. Unlatch the hooks of the J1 connector on the A10 PC board (Figure 4-28),
5. Remove the control panel from the upper G50 or the G60 S.
Page 30 of 52
hold the sub control panel in your hand.
sides of the flat-ribbon connector and disconnect the flat-ribbon cable from
the CN2 connector.
each transducer retainer and remove all of the retainers; the detent is
located on the bottom-inside edge and is removed downward.
Figure 4-25 Removing a Transducer Holder
Remove the six screws indicated in Figure 4-26.
Figure 4-26 Removing the Control Panel Fasteners
in Figure 4-27.
using the Phillips screwdriver, and disconnect the ribbon cable from J1.