Installing the Version 1.2 Software Update
1.0 Installing the Version 1.2 Software Update
Use the following instructions to install the version 1.2 software update on the
G60 S.

1.1 Preparing for the Version 1.2 Software Update

Use the following instructions to prepare for the installation of the version 1.2
software update.
The original release of the G60 S contained version 1.0 system software and
version 1.022 DIMAQ-IP software. The Speed Info 011/2002/US describes the
update of the version 1.022 DIMAQ-IP software to version 1.023.

1.1.1 Backing Up Service Data

Use the following steps to back up the service data.
1. Energize the G60 S.
2. Press the F4 key to display the Preset screen.
3. Select Service from the menu.
4. Enter the password and select the Ok button to display the service software
5. Select the Backup/Restore button to display the Backup / Restore screen.
6. Select the Backup radio button located at the top of the screen.
7. Check the All Packages checkbox.
8. Insert a formatted MO diskette in the MO drive and select the Go button
9. When the Format Disk message box is displayed, select the Cancel button
10. The Service Backup [file name] message box will be displayed; accept the
11. When the progress bar located on the Backup / Restore screen reaches
12. Select the Home button and the Exit button to exit the service software.
Page 2 of 8
Note: The original release of the G50 contains version 1.2 system software and
version 1.2 DIMAQ-IP software; the G50 does not require the installation of this
version 1.2 software update.
Note: On G60 S systems using the version 1.0 software, a formatted MO diskette
is required for the preset and service data backup operation because the MO
drive was incapable of formatting a diskette. Systems using version 1.2 software
support the formatting of MO diskettes.
Main Menu screen.
located at the bottom of the screen, to begin the backup.
to bypass the format operation.
backup name provided or enter a different name and select OK to continue
the backup operation.
100% a Done message will be displayed.