Chapter 5 - Installing Equipment Options

2.0 Installing the G50 Third Array Port

Use the instructions provided in this section to add a third array port to the G50
system.
!
2.1 Kit
Table 5-2 lists the contents of the kit provided for the installation of the
additional array port on the G50.
2.2 Tools Required
A number 2 Phillips screwdriver is required to perform this procedure.

2.3 Installing the Array Port

Use the following instructions to disassemble the G50 and install the additional
array port.
1. Verify that the system software is equal or greater than version 1.2.0; the
2. Move the G50 to a level location that provides 1 meter of clearance at the
3. Lock the 4 wheel locks.
4. De-energize the G50 and detach the power cord from the power outlet.
Siemens
Caution: The tasks described in this procedure expose static sensitive compo-
nents in the G50 to electrostatic discharge (ESD). Observe prescribed ESD pro-
cedures.
Part
A36 A3RD-2 PCB
M4 X 8 Screw
M3 X 8 Screw
Option Label
Note: In the following figures, fasteners are referenced as CK3 X 8 and CK4 X 8.
See "Fastener Reference" (page H-1) for information on these fastener designa-
tions.
additional port installed according to this procedure cannot be activated if
an earlier version of system software is installed on the G50. If the system
software requires upgrade to version 1.2.0, arrange to perform that task.
front and sides of the system.
Table 5-2 Kit Contents
Part Number or
Description
78 40 221
Binding head, external
lock washer
Binding head, external
lock washer
7482255 Rev 01
October 2002
Quantity
1
2
6
1
Page 11 of 16