PROline Prosonic Flow 93 FOUNDATION Fieldbus
Endress+Hauser
Parameter
Write access with
operating mode
(MODE_BLK)
DD_REV
read only
DEV_REV
read only
DEV_TYPE
read only
FAULT_STATE
read only
FEATURES
read only
Resource Block / base index 258
Description
Displays the revision number of the device description.
Note!
This display can be used to check whether the correct system files
(DD = Device Description) are being used for linking into the host
system.
The system files can be downloaded free of charge from the
Internet, (www.endress.com).
Example:
DD_REV parameter displays
DEV_REV parameter displays
Device description (DD) files required
Displays the revision number of the device.
Note!
This display can be used to check whether the correct system files
(DD = Device Description) are being used for linking into the host
system.
The system files can be downloaded free of charge from the
Internet, (www.endress.com).
Example:
DD_REV parameter displays
DEV_REV parameter displays
Device description (DD) files required
Displays the device type in decimal numerical format.
Display:
4158 (1059 hex) for Prosonic Flow 93 FOUNDATION Fieldbus
Current status display of the security behaviour of the Discrete
Output function block.
Display:
UNINITIALIZED
CLEAR (security behaviour not active)
ACTIVE (security behaviour active)
Displays the additional options supported by the device.
Display:
REPORTS
Event report is sent.
FAULTSTATE
The security behaviour of the Discrete Output function block is
supported, (see CLR_FSTATE and SET_FSTATE parameters).
HARD W LOCK
Hardware write protection
2 Resource Block
→
01
→
01
→
0101.sym and 0101.ffo
→
01
→
01
→
0101.sym and 0101.ffo
93
Need help?
Do you have a question about the PROline Prosonic Flow 93 and is the answer not in the manual?