Endress+Hauser PROline Prosonic Flow 93 Function Manual page 112

Foundation fieldbus ultrasonic flow measuring system
Hide thumbs Also See for PROline Prosonic Flow 93:
Table of Contents

Advertisement

3 Transducer Blocks (transmission blocks)
112
"TRANSDUCER_CH1" and "TRANSDUCER_CH2" (E+H parameters) /
Parameter
Write access
with operating
mode
(MODE_BLK)
System Unit -
AUTO - OOS
Viscosity
System Unit -
AUTO - OOS
Length
System Param. -
AUTO - OOS
Installation
Direction
Sensor
System Param. -
AUTO - OOS
Flow Damping
PROline Prosonic Flow 93 FOUNDATION Fieldbus
Base index 1200 and 1300
Description
For selecting the unit for the viscosity of the fluid.
Options:
mm²/s
cSt
St
Factory setting:
mm²/s
For selecting the unit for the measure of length.
The unit you select here is valid for:
• Nominal diameter
• Diameter
• Wall thickness
• Liner thickness
• Path length
• Wire length
• Sensor distance
Options:
MILLIMETER
INCH
Factory setting:
Depends on country (MILLIMETER or INCH),
see factory setting Page 76 ff..
This parameter can be used to reverse the sign of the flow rate
variable, if necessary.
Options:
NORMAL (FORWARD)
INVERSE (REVERSE)
Factory setting:
NORMAL (FORWARD)

Note!
System damping acts on all functions of the measuring device.
Specifies the filter depth of the digital filter.
This reduces the sensitivity of the measuring signal to interference
peaks (e.g. high solids content, gas bubbles in the fluid, etc.). The
measuring system reaction time increases with the filter setting.
The damping acts on all parameters and all downstream function
blocks.
User input:
0...100 s
Factory setting:
0 s
Endress+Hauser