Group Handling Totalizers - Endress+Hauser PROline Prosonic Flow 93 Function Manual

Foundation fieldbus ultrasonic flow measuring system
Hide thumbs Also See for PROline Prosonic Flow 93:
Table of Contents

Advertisement

6 Block TOTALIZERS
48
6.2

Group HANDLING TOTALIZERS

Þ
TOTALIZERS
D
TOTALIZERS
RESET ALL
Use this function to reset the totals
TOTALIZERS
(including all overflows) of the totalizers (1...3) to "0" (= RESET).
(3800)
Options:
NO
YES
Factory setting:
NO
FAILSAFE MODE
Use this function to define the common response of all totalizers (1...3) to error.
(3801)
Options:
STOP
The totalizer is paused until the fault is rectified.
ACTUAL VALUE
The totalizers continue to count based on the current measured flow value. The
fault is ignored.
HOLD VALUE
The totalizer continues to count the flow is based on the last valid flow value
(before the fault occurred).
Factory setting:
STOP
PROline Prosonic Flow 93 FOUNDATION Fieldbus
TOTALIZER 1
ß
TOTALIZER 2
ß
TOTALIZER 3
ß
HANDLING TOTALIZERS DJA
Function description
HANDLING TOTALIZERS
DAA
DAB
DAC
Þ
Handling totalizer functions
Handling totalizer functions
Endress+Hauser

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PROline Prosonic Flow 93 and is the answer not in the manual?

Table of Contents