[Procedure to key in adjustment values]
In accordance with the procedure described below, make adjustment of each adjustment item so that
the measured values obtained from test copies satisfy the specification. By pressing the [FAX] button,
immediately after starting the Adjustment Mode (05), single-sided test copying can be performed (nor-
mal copy mode).
<Operation keys>
0
5
POWER
1
Digital keys
: Enter codes.
START
Digital keys : Enter adjustment
values.
( CLEAR Use to make corrections)
ENTER
Value ZZZ is stored
replacing values YYY.
or
INTERRUPT
Test copy
FAX
If the test copy does not
satisfy the specified values,
return to step 1 and
repeat the adjustment
procedure.
Power OFF/ON : Exit the Adjustment Mode.
e-STUDIO200L/202L/203L/230/230L/232/233/280/282/283
ADJUSTMENT
<Display>
A
100%
TEST MODE
XXX
100%
TEST MODE
YYY
ZZZ
Current adjustment value
A
100%
TEST MODE
A
100%
Wait Warming Up
A
100%
COPYING
Fig. 3-6
© 2004 - 2010 TOSHIBA TEC CORPORATION All rights reserved
3 - 4
A3
Code No.
A3
Adjustment value newly entered
(If no entering, YYY is displayed.)
A3
A3
A3