Performance Verification

Battery Charge

Power-Up
Performance
Power-On Self-Test
10
Perform the following procedure to fully charge the battery.
1. Connect the monitor to an AC power source.
2. Verify that the monitor is off and that the AC Power/Battery Charging
indicator is lit.
3. Charge the battery for at least 14 hours with the monitor turned off or 18
hours with the monitor turned on.
The power-up performance tests verify the following monitor functions:
Power-On Self-Test, page 10
Power-On Defaults and Alarm Limit Ranges, page 11
1. Connect the monitor to an AC power source.
2. Verify that the monitor is off and that the AC Power/Battery Charging
indicator is lit.
3. Do not connect any cables to the monitor.
4. Turn on the N-595 by pressing the POWER ON/OFF button. Observe the
monitor front panel. The monitor must perform the following sequence:
Within 2 seconds all LEDs are illuminated, all pixels on the LCD
display are illuminated, and the backlight comes on.
The indicators remain lighted.
The LCD display shows NELLCOR and the software version of the
N-595.
Note: The software "Version" displayed in the example below is
X.X.X.X.The actual software version will be displayed on your
monitor.