Trend Data Printout (Ascii Mode) - Nellcor N-395 Operator's Manual

Table of Contents

Advertisement

Data Port Protocol
Table 4 1 : Status Codes
1 1
AS
I
Alarm Silence
1 1
Code
A 0
1 1
LB
I
Low Battery
Meaning
Alarin Off
1 1
LM
I
Loss of Pulse wl Motion
1 1
LP
Loss of Pulse
MO
Patient MOtion
1 1
PH
I
Pulse Rate High Limit Alarm
1 1
PL
I
Pulse Rate Low Limit Alarm
1 1
PS
I
Pulse Search
-
1 1
SH
I
Sat High Limit Alarm
1 1
1 1
SL
I
Sat Low Limit Alarm
1 1
1 1
SD
I
Sensor Disconnect
1 1
Note:
A sensor disconnect will also cause three dashes (-
-
-)
to
be displayed in the patient data section of the display or
printout.
TREND DATA PRINTOUT (ASCII MODE)
The fornat of data displayed when a trend printout (Figure 24) is
requested is similar to that of the real-time data. The only
differences are that "TREND" is displayed in the top row instead
of the "GRC: XXXX" software verification number and there is
no "Status" columa.
Readings are displayed in 4-second intervals. The values on each
row are an average for the 4-second period.
At the end of the printout an ""Output Complete" line indicates
that the transmission was successful. If the "Output Complete"
line is not present, a corruption of the data inay have been
detected and the data should be ignored.