Monitor Calibration - GE Vivid 3N Pro Series Service Manual

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
6-8-2

Monitor Calibration

Monitor calibration should be performed after adjusting the image quality settings using the monitor's
control buttons, as described in
1) Access the calibration options, as described in
2) Trackball to the Display option and press Select. The Calibration Display window is displayed, as
shown below:
3) Adjust the brightness of the monitor so that the black square in the upper-left corner of the window
is as dark as possible, while a faint border remains visible between that square and the square to
its right.
4) Rotate the soft key rotaries to offset the red, green and blue colors, to change the grey shades
displayed. By default, the RGB Calibration window shows one line at 45 degrees (Blue).
5) Trackball to the StopDisplayCalib button and press select to complete the monitor calibration.
04
EVISION
Figure 6-6 Calibration Display Window
Chapter 6 Service Adjustments
Section 6-6 "Monitor Operation" on page
Section 6-8 "Calibration" on page
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
6-10.
6-21.
M
ANUAL
6-23