Video Formats Confirmation

The Vivid 3N ultrasound scanner and VIC may be configured to operate with either PAL or NTSC video
systems, as required.
3-5-5-1
Video Format Confirmation
1) Turn ON the system.
2) Press Config.
3) From the System Configuration dialog box, select the VCR/ECG tab and make sure either PAL or
NTSC is selected.
4) Click the Technical Support tab and make sure Frequency is set to either 60Hz for NTSC, or 50Hz
for PAL.
5) Click OK.
Note:
If the video format is not set correctly, contact an authorized GE Service Representative.