3-6-2-4
Connecting the Color Printer
1) Place the color printer in the printer compartment above the B/W printer and VCR on the peripheral
tray, using an additional shelf.
2) Connect the following cables as shown in
Table 3-18 Color Printer Cables
From the Color Printer
NOTE:
All the color printer cables are located in the left storage compartment under the metal cover.
3) After connecting the remaining peripherals and switching the system ON, configure the printer
settings, as described in
3-6-2-5
Connecting the DeskJet Color Printer HP990Cxi
NOTE:
If there is sufficient space, the DeskJet Color Printer may be installed under the control console.
Alternatively, it will require a suitable stand or table to be positioned in close proximity to the
Vivid 3N scanner - at a distance of not more than 1m (3.3 ft) from the power connection to the
Vivid 3N unit.
WARNING: Whenever moving the Vivid 3N scanner, the DeskJet Color Printer must be
disconnected from the scanner. DO NOT attempt to move the two units simultaneously without
first disconnecting them. After relocation, re-connect the printer to the scanner.
1) Place the DeskJet color printer on the shelf below the control console (or alternatively on the
designated stand or table, adjacent to the unit).
2) Connect the following cables as shown in
Table 3-19 DeskJet Color Printer Cables
3) After connecting the remaining peripherals and switching the system ON, configure the printer
settings, as described in
For more details about peripherals installation refer to the Vivid 3N Peripherals Installation Manual.
3-26
04
EVISION
Video IN
External Trigger
AC IN
3-8-8 "Printers Tab" on page
From the Color Printer
Parallel Port Connector
AC IN (AC Dual Power cable)
3-8-8 "Printers Tab" on page
Section 3-6 - Completing the Hardware Installation
Table
3-18:
To the Cables
underneath the Control Console
Composite/Video OUT 1
Print Trigger
AC Power Cable
Table
3-19:
To the Right Panel
Parallel Port
Panel AC outlet