GE Vivid 3N Pro Series Service Manual page 252

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-3
Image Port (IMP) Diagnostic Tests
3) In the IMP tab, verify that the Comment area indicates that the system is ready to perform the test.
4) Trackball to the Self Test button and press Select to check the performance of the IMP board. The
Self Test LED changes color as follows:
5) Review the test results in the Comment area and proceed as follows:
7-14
04
EVISION
Yellow: The test is in progress.
Green: The test has been completed and no problems were identified.
Red: Problems were identified during the performance of the test.
If no problems were identified, proceed to the VME test, as described in
Diagnostic Tests( RFT Configuration)" on page
If no problems were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in
Section 6-8 "Calibration" on page
If errors were identified and the results were not marginal, verify that all image port and PCI
bridge cables are connected and functioning properly, as described in
and Function
(Theory). If the problem persists, contact your local GE Medical Systems field
engineer for assistance.
Section 7-3 - Performing Front End (FE) Diagnostics
V
IVID
(cont'd)
7-15.
6-21, and then repeat the IMP test.
3N P
/E
S
RO
XPERT
ERVICE
7-3-4 "VME Bus (VME)
Chapter 5 - Components
M
ANUAL