FK091075, R
IRECTION
3-9-6
Connecting Directly to EchoPAC PC
NOTE:
The EchoPAC PC workstation will reject any images that have a future date. Ensure that the
clock on the Vivid 3N Pro/Expert ultrasound unit lags behind that of the EchoPAC PC system
by a minute or two.
1) Press Config on the alphanumeric keyboard. The System Configuration window is displayed.
2) Trackball to the Connectivity tab and press Select. The Connectivity tab is displayed, with the
Views subsidiary tab selected by default, as shown in
3) Trackball to the Dataflow button and press Select. The full list of data flows is displayed, as shown
in
Figure 3-42 "Views Tab" on page
4) Expand the Remote Archive - Remote Storage branch. A service called Remote Servers is
displayed.
5) Expand the Remote Servers branch. A service called Remote Archive - Remote Storage is
displayed.
6) Trackball to the Services tab and press Select. The Services tab is displayed, as shown in
Figure 3-39 "Services Tab" on page
7) Select the EchoPAC server from the Name dropdown list. The Services tab is updated, as shown
below:
8) Enter the IP address of the EchoPAC server in the IP-address field.
3-62
04
EVISION
Figure 3-45 Services Tab - EchoPAC
Section 3-9 - Connectivity Setup
V
Figure 3-36 "Connectivity Tab" on page
3-58.
3-56.
3N P