GE Vivid 3N Pro Series Service Manual page 301

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-9-6-2 AC Power Input Test
The AC power input test measures the AC input voltage deviation from the nominal input voltage
(accuracy by percentage).
1) Access the MUX tab of the FB Diagnostics dialog box, as described in
Diagnostic Test Options" on page
2) Trackball to the Advanced board tests button and press Select. The Advanced System Tests
options are displayed, as shown in
61.
3) Trackball to the Monitor button and press Select. The View Power options are displayed, as shown
below:
04
EVISION
Figure 7-37 View Power Options
Chapter 7 Diagnostics/Troubleshooting
7-50.
Figure 7-35 "MUX Advanced System Tests Options" on page 7-
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
7-3-9-1 "Accessing the MUX
M
ANUAL
7-63