GE Vivid 3N Pro Series Service Manual page 141

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
4-4-1-5
CFM Stationary Noise Test
Table 4-22
Step
1
2
3
4
4-4-1-6
Probe Operation Check (Alignment and Sensitivity) Test
Table 4-23
Step
1
2
04
EVISION
CFM Stationary Noise Test (3S)
Task
Use the vertical nail area in the Electronic
Flow Phantom RMI1425A.
Place the 3S probe above the vertical nails
line (position 1).
Select a Cardiac preset and activate CFM.
Set the following CFM parameters:
Gain: Max possible without color noise
Tissue Priority: 0
Flash: 0
Probe Operation Check (Alignment and Sensitivity) Test (3S)
Task
Use the Electronic Flow Phantom RMI1425A.
Set the Phantom settings as follows:
Preset: 3
Mode: Steady
Chapter 4 Functional Checks
V
3N P
IVID
Expected Result(s)
Ensure that the color does not appear constantly on the white nail.
Expected Result(s)
/E
S
M
RO
XPERT
ERVICE
ANUAL
4-13