GE Vivid 3N Pro Series Service Manual page 305

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-9-7-2 Temp Test
The Temp test displays the temperature at the Front End crate, as well as the temperature of the AC
distribution box.
1) Access the MUX tab of the FB Diagnostics dialog box, as described in
Diagnostic Test Options" on page
2) Trackball to the Advanced board tests button and press Select. The Advanced System Tests
options are displayed, as shown in
61.
3) Trackball to the Monitor button and press Select. The View Power options are displayed, as shown
in
Figure 7-37 "View Power Options" on page
4) Trackball to the Temperature button and press Select. The temperatures are displayed, as shown
below:
5) Trackball to the Show button and press Select to display the temperatures.
NOTE:
For repeat readings, select Loop every and set the reading time cycle.
04
EVISION
Figure 7-41 Temperatures
Chapter 7 Diagnostics/Troubleshooting
7-50.
Figure 7-35 "MUX Advanced System Tests Options" on page 7-
7-63.
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
7-3-9-1 "Accessing the MUX
M
ANUAL
7-67