GE Vivid 3N Pro Series Service Manual page 355

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-5-8-2
Scanner Utilities (cont'd)
Clean UserDefs-used to clean all user-defined settings and restore the default settings
CAUTION: THIS PROCESS IS IRREVERSIBLE! Before proceeding, make sure you have a
complete backup of the current settings if they will be required in the future.
When selecting Clean Userdefs, the following warning message is displayed:
After making a back-up copy of the current settings, click OK to proceed.
NOTE:
This function performs the same procedure as that of Restore Factory Presets (accessed under the
System Configuration window - System tab).
04
EVISION
Figure 7-90 Clear User-defined Settings
Chapter 7 Diagnostics/Troubleshooting
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
M
ANUAL
7-117