BEP2 - ECG Module Removal Procedure
NOTE:
BEP2 refers to a Pentium 4-based BEP (newer type).
1) Disconnect the mains AC power cable for the mains power wall outlet.
2) Remove the Vivid 3N left and right side covers - refer to the
Procedure" on page
3.) Remove the Vivid 3N rear cover - refer to the
7.
4) Disconnect all cables from the rear of the BEP.
5) Remove the BEP from the machine as described
Procedure" on page
6) Remove both BEP side covers as described in
on page
7) Remove the BEP front cover as described in
page
in.Figure 8-71
8) Remove the four screws that secure the ECG Module in position (two from each side of the exposed
BEP as shown in
8-84
04
EVISION
8-4.
8-57.
8-64.
8-65. A left view of the exposed BEP showing the location of the ECG module is seen
below:
Figure 8-71 ECG Module - BEP2
Figure 8-72 "ECG Module Removal- BEP2" on page
Section 8-5 - Back End Parts Replacement
V
3N P
IVID
8-2-2-3 "Side Covers Removal
8-2-4-3 "Rear Cover Removal Procedure" on page 8-
8-5-1 "Back End Processor Replacement
8-5-3-3 "BEP2 - Side Cover Removal Procedure"
8-5-3-5 "BEP2 - Front Cover Removal Procedure" on
/E
S
M
RO
XPERT
ERVICE
ANUAL
ECG Module
8-85).