GE Vivid 3N Pro Series Service Manual page 303

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-9-7
LVPS Voltage Test
The LVPS test measures the LVPS voltages. The voltage range is displayed next to the measured
voltage. All the voltages can be measured with DVM on the front board test point.
1) Access the MUX tab of the FB Diagnostics dialog box, as described in
Diagnostic Test Options" on page
2) Trackball to the Advanced board tests button and press Select. The Advanced System Tests
options are displayed, as shown in
61.
3) Trackball to the Monitor button and press Select. The View Power options are displayed, as shown
in
Figure 7-37 "View Power Options" on page
4) Trackball to the LV PS Voltage button and press Select. The LVPS voltage ranges are displayed,
as shown below:
5) Trackball to the Show button and press Select to display the voltages.
NOTE:
For repeat readings, select Loop every and set the reading time cycle.
04
EVISION
Figure 7-39 LVPS Voltage Ranges
Chapter 7 Diagnostics/Troubleshooting
7-50.
Figure 7-35 "MUX Advanced System Tests Options" on page 7-
7-63.
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
7-3-9-1 "Accessing the MUX
M
ANUAL
7-65

Advertisement

Table of Contents
loading

Table of Contents