Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
7-3-8-4
Receive Test (cont'd)
The Report dialog box contains the following information:
Test Result Message: Displays the test result.
TR4 Modules: Displays a graphic representation of the front board (four groups of TR4 sockets).
Each group consists of four TR4 sockets, each socket containing four vectors (channels), as
follows:
-
-
-
Checkboxes: Enable you to manipulate bad channels:
-
-
04
EVISION
Figure 7-21 Report Dialog Box - Receive Test Low Frequency
Each vector is represented by a button (highlighted in green) that displays the number of the
socket to which the TR4 is connected.
Each button is highlighted yellow when selected.
Each button is highlighted red if errors were detected for one or more channels.
Show Only Bad Channels: Select this checkbox to display only bad channels in the Display
area.
Mark bad channel: Select this checkbox to manually mark the currently selected channel as
bad.
Chapter 7 Diagnostics/Troubleshooting
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
M
ANUAL
7-39