GE Vivid 3N Pro Series Service Manual page 503

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
8-8-3-7
JVC VCR Removal Procedure
1.) Remove the Connectors Panel cover as shown in
Procedure" on page
2.) Disconnect the RS-232C cable, the Audio and S-Video IN and OUT cables, and the Power cable
from the Vivid 3N rear connectors panel - see
on page
3.) Remove the adhesive velcro strip from the edge of the side compartment on the left side of the Vivid
3N ultrasound unit (refer to
4.) Slide the VCR partially out of the compartment, carefully easing the attached cables with it.
5.) Disconnect the RS-232C cable, the Audio and S-Video IN and OUT cables from VCR rear
connectors panel (refer to
6.) Disconnect the AC power cable from the AC Input socket on the VCR rear panel (upper left corner).
7.) When all cables are disconnected, remove the VCR completely from the Vivid 3N ultrasound unit.
04
EVISION
8-8.
8-132.
Figure 8-113 "JVC VCR - Rear Connectors Panel" on page
Chapter 8 Replacement Procedures
8-2-5-3 "Connector Panels Cover Removal
Figure 8-115 "Vivid 3N Left Rear Panel Connectors"
Figure 8-114 "JVC VCR Installed in Compartment" on page
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
M
ANUAL
8-131).
8-130).
8-133