GE Vivid 3N Pro Series Service Manual page 402

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
8-3-2-4
Vivid 3N 17" Monitor Installation - Procedure1
NOTE:
Before mounting the new monitor into position, make sure that the Teflon washer has
remained in place on top of the monitor mounting base.
1.) In the upright position, carefully lift the new monitor and insert its pivot shaft (see
through the Teflon washer and into the hole on the monitor base.
While lowering the monitor downwards, swivel it slightly to the right and left to ease insertion.
2.) Check that the pivot shaft is fully inserted, and that the monitor base is resting on the Teflon washer.
3.) Position the monitor directly in line with the front of the control console and tighten the securing
screw located at the rear of the monitor base.
4.) Mount the plastic cover of the base assembly back into position and secure with the four fastening
screws provided (two on each side).
5.) Plug the VGA cable and AC power cable into the rear of the control console.
6.) Check that the monitor is sufficiently loose to enable swivelling slightly from left to right.
7.) Make sure the monitor can be tilted up and down, and that it maintains its set position.
8.) Adhere the appropriate Vivid 3N system label.
For the next stage of the new monitor installation procedure, proceed to the instructions for
Quality Calibration for the Vivid 3N 17" Monitor" on page
8-32
04
EVISION
Figure 8-33 New Samsung 17" Monitor - Viewed from Underside