Monitor Calibration

Monitor calibration should be performed after adjusting the image quality settings using the monitor's
control buttons, as described in the
1) Access the calibration options, as described in the
2) Trackball to the Display option and press Select. The Calibration Display window is displayed, as
shown below:
3) Adjust the brightness of the monitor so that the black square in the upper-left corner of the window
is as dark as possible, while a faint border remains visible between that square and the square to
its right.
4) Rotate the soft key rotaries to offset the red, green and blue colors, to change the grey shades
displayed. By default, the RGB Calibration window shows one line at 45 degrees (Blue).
5) Trackball to the StopDisplayCalib button and press select to complete the monitor calibration.