GE Vivid 3N Pro Series Service Manual page 174

Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
FK091075, R
IRECTION
5-4-3-5
Transmit Signal Path
The transmit signal is configured via the operating mode that was selected by the user. When this mode
is selected, it sets those operating parameters for the FEC, and the FEC sends a signal to the high
voltage power supply (HVPS) to program one of the supplies (HVPS1 or HVPS2).
Once the signal is transmitted, it is received by the transmit pulsor (TP) which is a switching device that
will propagate the signal between different elements of the probe. The propagation of that signal
depends on the way in which the FEC programs the Front board RAM table.
The signal from the HVPS comes in to the TP; the TP then generates a signal to the probe (connected
to the Front board), in accordance with a command generated in Timing Pulse Generator (TPG). The
TPG operates according to tables in the RAM table - see
5-12
04
EVISION
Front Board
Front Board
Figure 5-6 Transmit Signal Path Block Diagram
Section 5-4 - Front End
V
3N P
/E
IVID
RO
XPERT
Figure 5-6
on page 5-12.
S
M
ERVICE
ANUAL