Pencil Probe Test (cont'd)
The Report dialog box includes the following:
•
•
•
5) Check the test result message and proceed as follows:
•
•
6) Trackball to the Start Close test and press Select. The pencil probe test (close) is performed and
the test results are displayed in the Report dialog box, as shown in
- Pencil Probe Test (Close)" on page
7-44
04
EVISION
Test Result Message: Displays the test result.
Pencil Probe Relays: Displays a graphic representation of the eight front board pencil probe
relays, as follows:
•
Each relay is represented by a button (highlighted in green) that displays the relay
number.
•
Each button is highlighted in yellow when selected.
•
Each button is highlighted in red if errors were detected for one or more relays.
Checkboxes: Enable you to manipulate bad channels:
•
Show Only Bad Channels: Select this checkbox to display only the bad channels in the
Display area.
•
Mark bad Channel: Select this checkbox to manually mark the currently selected channel
as bad.
If one or more relays were identified as bad, trackball to Save Report and press Select, and
contact a GE Medical Systems service representative.
If no relays were identified as bad, mark the channels suspected as being bad as follows:
•
Trackball to the channel in the Display area and press Select.
•
Trackball to the Mark bad channel checkbox and press Select. Repeat as required to
mark additional channels.
•
Trackball to Save Report and press Select, and contact a GE Medical Systems service
representative.