Service Log In
1) From the System Configuration window, make sure that iLinq (the Common Service platform) is
enabled under the Technical Support tab, as shown in
RFI-configured systems
RFT-configured systems
2) Trackball to OK and press Select. The System Configuration window closes.
NOTE:
If iLinq was not found to be enabled in
proceeding.
3) Trackball to the wrench icon
Select.
The Profile Manager window is displayed, as shown in
page
04
EVISION
Figure 7-59 Technical Support - System Configuration
7-90.
Chapter 7 Diagnostics/Troubleshooting
Figure 7-59
step
1, it is necessary to re-boot the system before
in the status bar at the bottom of the scanning screen and press
Figure 7-60 "Profile Manager Window" on
V
3N P
/E
S
IVID
RO
XPERT
ERVICE
below.
M
ANUAL
7-89