Validating Communication with the EchoPAC Server
1) Trackball to the Views tab and press Select. The Views tab is displayed.
2) Trackball to the Dataflow button and press Select. The full list of data flows is displayed, as shown
below.
3) Expand the Remote Archive - Remote Storage branch. A service called Remote Servers is
displayed.
4) Expand the Remote Servers branch. A service called Remote Archive-Remote Storage is
displayed.
5) Trackball to Remote Servers and press Select, then trackball to the Check button and press
Select.
6) If the connection was successfully established, trackball to the Remote Archive-Remote Storage
line and press Select, then trackball to the Check button and press Select.
04
EVISION
Figure 3-48 Views Tab
If the connection is established, a green checkmark is displayed.
If the connection cannot be established, a red cross is displayed.
If the connection is established, a green checkmark is displayed.
If the connection cannot be established, a red cross is displayed.