Validating Communication with the Workstation (cont'd)
5) Press <Ctrl+S> to view the status of the spooler, as shown below:
The Status column can contain the following options:
04
EVISION
Figure 3-44 Job Spooler
Hold:
Images are in the spooler buffer but transmission has not begun as the exam is
active.
Active:
Images are being transmitted over the network.
Done:
Transmission is complete and all the images have been transmitted over the
network.
Fail:
Transmission failed or is incomplete. An error message will be displayed in the Error
column.
Chapter 3 Installation