Top Cover Lower Section Removal Procedure (cont'd)
9) Remove the section of the collar released in the previous step, leaving an open gap at the rear of
the top cover, as shown below:
10) Remove the four screws that secure each connector panel to the top cover.
11) Gently pull the rear panels backwards, leaving them hanging loosely on their cables, as shown
below:
12) Remove the two screws securing the rear section of the gas spring cover, and then remove this
cover.
04
EVISION
Figure 8-12 Top Cover with Collar Removed
Figure 8-13 Connector Panels - Loosened
Chapter 8 Replacement Procedures