IRECTION
10-3-1
How Often Should PM Procedures be Performed?
Service at Indicated Time
Clean Probes
Clean Probe Holders
Clean Air Filter
Inspect AC Mains Cable
Inspect Cables and Connectors
Clean Console
Clean Monitor
Inspect Wheels, Casters, Brakes and
Swivel Locks
Check Control Console Movement
Console
Current Leakage Checks
Peripheral
Current Leakage Checks
Surface Probe
Current Leakage Checks
Endocavity Probe (vaginal or anal)
Current Leakage Checks
Transesphongeal (TEE) Probe
Current Leakage Checks
Surgical Probe
Current Leakage Checks
Measurement Accuracy Checks
Probe/Phantom Checks
Functional Checks
10-4
04
EVISION
Table 10-2 Periodic Maintenance Schedule
Daily
Weekly
•*
*
*
Section 10-3 - Periodic Maintenance Schedule
Monthly
Annually
*
V
3N P
/E
IVID
RO
XPERT
(cont'd)
Notes
* or before each use
More frequently if necessary, depending on
the environment
*Mobile Unit Check Weekly
* Mobile Unit Check Daily
* Mobile Unit Check Daily
Also after corrective maintenance,
or as required by your facility's QA program.
Also after corrective maintenance,
or as required by your facility's QA program.
Also after corrective maintenance,
or as required by your facility's QA program.
As prescribed in Probe Manual.
As prescribed in Probe Manual.
As prescribed in Probe Manual.
Also after corrective maintenance,
or as required by your facility's QA program.
Also after corrective maintenance,
or as required by your facility's QA program.
Also after corrective maintenance,
or as required by your facility's QA program.
S