Front Cover Removal Procedure (cont'd)
4) Remove the three screws located across the front of the top cover, shown below:
5) In order to free the front cover, grip the cover by its lower sides, and push upwards while pulling it
towards you.
8-2-3-4
Front Cover Installation Procedure
1) Return the front cover to its original position at the front of the unit, taking care to first insert its upper
edge under the top cover.
2) Fasten the front cover with the three screws at the edge of the top cover, as shown in
above.
3) Fasten the left and right brackets at the lower end of the front cover, one each side, by tightening
each bracket with its two screws to the main metal frame, as shown in
Bracket Screws" on page
4) Install the side covers, as described
5) Insert the air filter to its original location, as shown in
NOTE:
It is recommended to clean the filter before returning it to its original location, as
described in
8-6