Drystar 5500, Drystar 5503 | Advanced Operation (key-operator mode) | 63

Print head profile calibration

To start the manual print head profile calibration:
1. Press the Down key once to select 'Printhead profile', and press the
Confirm key.
The 'Select input tray' screen appears:
2. Press the Up/Down arrow keys to select an input tray, followed by the
Confirm key (e.g. choose upper input tray).
The system will start with Phase 1 of the calibration: Registration:
As a result, the 'Registration' print out appears.
3. After registration, the system will continue with Phase 2 of the calibration:
the print head profile calibration:
As a result, the 'TPH profile' print out appears.
4. In case the print head profile calibration has failed, the following message
appears:
WARNING:
Print head profile calibration must be done whenever the profile
of the print head needs to be improved (manual print head
profile calibration).
Note: In the procedure, we will assume a print head profile
calibration on the upper input tray. The procedure for the lower
input tray is identical.